Tektronix TDS3000(B) Rotary Encoder Repair
- Introduction
- Opening Up the Oscilloscope
- Cleaning the Bourns Rotary Encoders
- Detented and Non-Detented Rotary Encoders
- Reinstallation
- Success!
- References
Introduction
One of the common issues of TDS3000(B) oscilloscopes is that their rotary knobs fail after a while. In addition to missing knobs and a broken shaft for the Horizontal Scale knob, the top-left general purpose knob of my TDS3014B behaved erratically.

Their rotary encoders were made by Bourns, created specifically for Tektronix and are not available on the open market.
Fortunately, fixing them is possible without replacement parts. In this blog post, I show how to do it.
Opening Up the Oscilloscope
In my TDS3000 LCD Repair blog post, I already describe in detail how to open the scope. Just follow those instructions until you reach the point where the front panel assembly is loose:

You need to free the front panel PCB from the plastic enclosure:
-
Carefully remove the bluish flex cable.
I used needle-nose pliers for this.
-
Unclip the 6 plastic clips.
They’re pretty stiff. I used a screwdriver to push against them. I was able to get the panel loose without breaking any, but on another unit someone else had already given it a try and 2 clips were broken.
-
Cut away the center melted plastic blob.
On my scope, there was a deformed (melted?) piece of plastic that made it impossible to remove the PCB without cutting it away with a micro shear flush cutter. See the red rectangle in the picture below.
However, in that other scope, there was a regular alignment post sticking through, like a bunch of others, and nothing had to be cut.

Cleaning the Bourns Rotary Encoders
The rotary encoders consist of 3 parts:
- A plastic case
- A plastic rotor with three-finger wiper contact
- PCB contact rings

The plastic case is attached to the PCB with heat staking: the 4 plastic studs go through the PCB and are then melted and deformed, a bit like rivets:

This is a one-way process, unfortunately: the only way to remove the encoder case is by cutting away those black heat-staked heads. To reinstall them, you have different options. We’ll get to that.
Either way, it’s clear that there isn’t much that can go wrong with the encoder itself: a failure is almost certainly due to a bad contact between the wiper contacts and the PCB.
After removing the encoder, I sprayed a bit of DeoxIT on the PCB, temporarily reinstalled the case and rotor, and gave it a bunch of turns to make sure the contacts got some DeoxIT on them as well. I then cleaned up the DeoxIT.
Detented and Non-Detented Rotary Encoders
In addition to a failing knob, the shaft of the Horizontal Scale knob had broken off. Lacking a donor board for a spare, I decided to swap the rotary encoders of the Horizontal Scale and the Waveform Intensity knobs; I rarely use the latter.
The Vertical Scale and Horizontal Scale knobs use detented rotary encoders: they have a discrete feel to them instead of a smooth rotation. This is achieved with a spring between the plastic case and the rotor, they’re otherwise identical.

If you’re swapping or fixing multiple rotors, make sure to install the spring in the correct ones!
Reinstallation
There are different ways to fix the encoder case back in place. Epoxy, glue, and soldering iron heat are suggested solutions on the EEVblog forum. I used simple tape instead. Because here’s the thing: once put back together, the rotary encoders are squeezed tightly between the plastic front panel and the PCB; there is no way they’ll come loose. The tape is only there to keep the encoder in place during reassembly.

Success!
I’ve become very good at getting everything back together. Once done, the broken rotary knob is working fine again. The Horizontal Scale knob is still working too, with an intact shaft this time; even the Wave Intensity rotor still works though you need a screwdriver to do so.
I transplanted the knobs of my TDS3052 to the TDS3014B: